Abstract

Analog to digitalconverter circuit design for biomedical systems with multiple recording channels presents challenges in high density and very low power consumption. Passive integrator and loop-filter based delta-sigma modulators (DSMs) have been recently reported for ultra-low-power and highly energy-efficient data converters for multi-channel biopotential acquisition. However, these modulators rely on a very high oversampling ratio (OSR) to achieve the target resolution. Higher OSR leads to higher power consumption in the modulator and the digital low-pass and decimation filter succeeding the DSM. We present a low OSR passive integrator-based DSM in this work by relying on a duty-cycled resistor (DCR). DCR enables the realization of large time constants in the already passive loop-filter, with minimal area and overhead power consumption. This leads to design of DSMs that are highly area, power and energy-efficient, suitable for multi-channel biopotential recording systems. We demonstrate a second order, duty-cycled passive integrator based CTDSM in a 65 nm CMOS technology for a 10 kHz biopotential bandwidth. Measurement results show that the fabricated design achieves an SNDR/DR of 56.36/63.1 dB while consuming only 160 nW power with an OSR of 32 and occupies an area of 0.035 mm 2 with a state-of-the-art energy efficiency of 14.9 fJ/conv. In-vitro and in-vivo measurements are provided to further demonstrate the operation of the proposed DSM.