We have evaluated two automated methods for measuring ristocetin cofactor activity (VWF:RCo) on an automated coagulation analyzer (STAR, Diagnostica Stago Inc., Parsippany, NJ). A modification of the BC von Willebrand Reagent method (Dade Behring Inc., Newark, Delaware) was compared with a standard method using a platelet aggregometer, and an in-house automated method was developed using commercially available components. The STAR measures the change in optical density (OD) caused by agglutination of platelets. Change in OD was linear for plasma dilutions between .25 and .04 U/mL. Coefficient of variation compared favorably with that of the standard method. In 123 normal women, the correlation coefficient (r) = .905. In menorrhagia cases, r = .901. In 20 VWD patients, r = .93. The CDC in-house method was compared to the BC Reagent method in 79 additional menorrhagia cases, with r = .94. The automated methods produced results equivalent or superior to those of traditional methods of measuring VWF:RCo.