Preservation of the breath sample integrity during storage and transport is one of the biggest challenges in off-line exhaled breath gas analysis. In this context, adsorbent tubes are frequently used as storage containers for use with analytical methods employing gas chromatography with mass spectrometric detection. The key objective of this short communication is to provide data on the recovery of selected breath volatiles from Tenax® TA adsorbent tubes that were stored at −80 °C for up to 90 d. For this purpose, an Owlstone Medical’s ReCIVA® Breath Sampler was used for exhaled breath collection. The following fifteen compounds, selected to cover a range of chemical properties, were monitored for their stability: isoprene, n-heptane, n-nonane, toluene, p-cymene, allyl methyl sulfide, 1-(methylthio)-propane, 1-(methylthio)-1-propene, α-pinene, DL-limonene, β-pinene, γ-terpinene, 2-pentanone, acetoin and 2,3 butanedione. All compounds, but one (acetoin), were found to be stable during the first 4 weeks of storage (recovery within ± 2 × RSD). Furthermore, n-nonane was stable during the whole of the investigated period.
Read full abstract