Dissociative electron attachment (DEA) of a molecular target XY, e- + XY → XY- → X + Y-, is an important process in plasma, atmosphere, interstellar space, and ionizing radiation. DEA dynamics, i.e., the formation and fragmentation of an electron-molecule resonant complex, can be unveiled by measuring the product Y- with the velocity map imaging (VMI) technique. However, it is still challenging to achieve a high-resolution VMI measurement. Recently, we developed a high-resolution DEA apparatus that combined the VMI technique with a trochoidal electron monochromator. The energy spread (around 500meV) of thermally emitted incident electrons can be reduced remarkably, but the monochromatized electrons become diffuse again in energy when being pulsed with an electric pulse applied on the grid electrode because the electron beam must be pulsed in the VMI measurement. Now this long-standing technical contradiction is settled by introducing a parallel resistance-capacitor circuit to the electrode of an electron gun. A delay-line detector is used for the VMI measurement, which allows the three-dimensional velocity or momentum images of multiple anionic yields to be recorded efficiently. Our high-resolution VMI apparatus works at a pulse frequency of 5kHz and an energy spread of about 120-150meV, and its credible performances are exhibited by the measurements of the DEA processes of CO(a3Π) and NO2.