Air-Displacement-Plethysmography (ADP) by BOD POD is widely used for body fat assessment in children. Although validated in healthy subjects, studies about use in pediatric patients are lacking. We evaluated user experience and usability of ADP measurements with the BOD POD system in healthy children and pediatric and young adult patients. Using the experiences of seven cohort studies, which included healthy children and patients aged 2-22 years, we retrospectively evaluated the user experience with the User Experience Questionnaire (UEQ) (n=13) and interviews (n=7). Technical performance was studied using the quality control data collected by the ADP-system. From 2016 to 2022, 1606 measurements were scheduled. BOD POD was mostly rated 'user-friendly', with a generally neutral evaluation on all scales of the UEQ. However, questionable reliability and validity of the results were frequently (86%) reported. We found a high technical failure-rate of the device, predominantly in stability (17%) and accuracy of the measurement (12%), especially in the 'pediatric option' for children aged <6 years. Measurement failure-rate was 38%, mostly due to subject's fear or device failure, especially in young and lean children, and in children with physical and/or intellectual disabilities. We conclude that ADP by BOD POD in children and young adults is non-invasive and user-friendly. However, in specific pediatric populations, BOD POD has several limitations and high (technical) failure-rates, especially in young children with aberrant body composition. We recommend caution when interpreting body composition results of pediatric patients as assessed with BOD POD using the current default settings.