BackgroundSafely implementing transbronchial lung cryobiopsy (TBLC) in patients with interstitial lung disease (ILD) requires accurate navigation. Traditional fluoroscopy falls short in reducing the risk of post-procedure pneumothorax. The potential of electromagnetic navigation bronchoscopy (ENB) as a more precise navigation method warrants further exploration.MethodsA prospective cohort study was conducted on ILD patients undergoing TBLC. Patients were assigned either fluoroscopy or ENB for cryoprobe positioning. Navigation accuracy was evaluated using cone beam computed tomography (CBCT) images as the standard. Safety and diagnostic yield were also observed.ResultsSeventeen patients underwent TBLC, with 10 guided by fluoroscopy and seven by ENB. Fluoroscopy-guided cryoprobe navigation required more adjustments [9/15 (60%) v.s. 1/9 (11%), p = 0.018] for subsequent TBLC compared to ENB, as confirmed by CBCT images. Clinical characteristics, post-procedure complications, and biopsy specimen size showed no significant differences between the groups. Fourteen patients obtained a pathological diagnosis, and 15 received a multidisciplinary discussion (MDD) diagnosis. In the fluoroscopy group, three patients failed to obtain a pathological diagnosis, and two failed to obtain an MDD diagnosis.ConclusionsENB demonstrates significantly superior accuracy in TBLC navigation compared to traditional fluoroscopy when CBCT images are used as a reference. Further studies are necessary to determine the value of ENB in TBLC navigation for ILD patients.