Purpose: To establish a method of quality assurance for a novel aperture modulated translating bed total body irradiation technique currently under development at the Tom Baker Cancer Centre. Methods: This TBI technique uses the MLC on a conventional linear accelerator to dynamically shape the radiation beam as the patient is translated through with 0° gantry angle. The dynamically shaped field allows for the compensation of varying patient composition as the patient translates through the beam, resulting in improved dose homogeneity. This technique requires a high degree of accuracy in the positioning of the translating bed and MLC leaves throughout treatment, as the bed position must be properly correlated with the field shape being delivered. The QA method developed in the present work makes use of the linacˈs EPID to verify the location of the individual leaves of the collimating system throughout treatment delivery. The EPID operates in cine mode, and acquires images of the MLC defined field throughout plan delivery. The MLC defined field edges of each acquired image are found using a simple edge detection algorithm, and the locations of the detected edges are compared to the prescribed MLC positions for the specified number of MUs delivered. The position of the translating bed is also monitored throughout the delivery to ensure proper correlation between MLC leaf positions and bed position. Results: The EPID allows for the accurate detection of bed and MLC leaf positions throughout the delivery of aperture modulated translating bed TBI plans. Quantitative data comparing prescribed and detected positions of the bed and MLC leaves are supplied for the entire duration of plan delivery. Conclusions: The method developed allows for pre‐treatment verification of the novel TBI treatment plans, and has the potential to be extended in order to verify MLC and bed positions during actual patient treatment.