To evaluate postoperative masticatory efficacy of a new design non-compression titanium miniplate compared to conventional non-compression titanium miniplate on the basis of bite force for treatment of mandibular angle fractures. The prospective study included 20 patients with mandibular angle fractures randomly categorized into 2 groups: Group I, fixation of angle fractures by conventional miniplates, and Group II, fixation of angle fractures by new design miniplates. Evaluation was done for clinical outcome, primarily bite force; radiological outcome; and associated postoperative morbidities at different time intervals. The results showed to be highly significant in terms of mean operating time for plate adaptation and fixation and bite force adaptation (p = 0.003 at follow-up of 6 months) for the newer miniplate compared to the conventional miniplate. No statistically significant difference was seen for postoperative paresthesia, malunion, non-union, occlusal discrepancy, or hardware failure. Within the limits of the study, it appears that the single, monocortical, non-compression, superior border new design miniplate proved to be a successful procedure for treating non-comminuted mandibular angle fractures specifically in terms of enhanced postoperative masticatory efficiency as compared to conventional miniplates. Further clinical studies with larger sample size can derive a more comprehensive conclusion.