BackgroundIt is unclear whether the use of polyetheretherketone (PEEK) interference screws for anterior cruciate ligament (ACL) reconstruction leads to postoperative tunnel enlargement. Femoral tunnel enlargement was further compared with adjustable-length loop cortical fixation.MethodsEighteen patients with ACL reconstruction using hamstring grafts were retrospectively divided into two groups. Eleven patients were treated with the ACL reconstruction technique using a PEEK interference screw for femoral graft fixation. Seven patients received adjustable-length loop cortical buttons for femoral fixation. Tibial ACL graft fixation was performed using PEEK interference screws. Tunnel volume changes were assessed using computed tomography (CT) scans performed after surgery (100%) and after 1 year. The maximal tunnel diameter was measured.ResultsThe group with femoral screw fixation showed a mean tunnel volume change of 108.15 ± 13.7% on the tibial side and 124.07 ± 25.38% on the femoral side. The group with femoral button fixation showed a tunnel volume change of 111.12 ± 12.72% on the tibial side and 130.96 ± 21.71% on the femoral side. The differences in femoral tunnel volume changes were not significant (P = 0.562). Femoral tunnels with PEEK screw fixation showed significantly larger diameter after 12 months in comparison with button fixation (13.02 ± 1.43 mm vs. 10.46 ± 1.29 mm, P < 0.001).ConclusionsPEEK interference screws were associated with significant tibial and femoral tunnel enlargement. Femoral tunnel enlargement was comparable between PEEK interference screws and button fixation. Final femoral tunnel diameter was significantly larger with PEEK screw fixation in comparison to button fixation.