Laser ultrasonics is a noncontact measurement method that uses a laser-induced elastic wave source in combination with an optical surface displacement-tracking system. This study compared the performances of two optical interferometers with different characteristics when applied to measurement of pulsed thermoelastic waves. The surface displacement-tracking system was designed to measure the center of the microscopic view. A pulsed laser beam irradiated a black ink layer to generate the thermoelastic waves. The out-of-plane displacement on the axially opposite side was then measured using either a Michelson interferometer or a Sagnac interferometer. The objective lens of the system was of a type commonly used in biological observations. The Michelson interferometer estimated a maximum displacement of 0.43 nm and a maximum sound pressure of 24.7 kPa. The signal-to-noise ratios from 16 averages were 14.9 dB (Michelson interferometer) and 19.2 dB (Sagnac interferometer). Furthermore, this paper compares the performance of the numerically estimated Sagnac interferometer outputs calculated from the measured Michelson interferometer outputs with the experimentally obtained Sagnac interferometer outputs. The numerically estimated Sagnac interferometer’s output was shown to be identical to the experimentally acquired output. The Michelson interferometer requires a higher average operating frequency (i.e., it needs a longer data acquisition time), although this interferometer does offer superior displacement output linearity. This property enables calculation of the sound pressure from the displacement amplitude. These findings indicated that combination of the measurement points of the Sagnac interferometer with those of the sparsely distributed Michelson interferometer reduced the measurement time when compared with a single use of the Michelson interferometer while also maintaining the data acquisition quality.
Read full abstract