Heart valve disease is commonly treated by minimally invasive procedures with guide wires and catheterization. The main purpose of this study is to find out whether an extension of the guide wire with a sensor can support the surgeon within the blood vessel to reduce X-ray necessity. A smart guide wire is developed by an extension with a flex-bending sensor to evaluate the sensor signal with and without "blood" flow at a constant compression force. Various surgically relevant investigations are performed. For assessment, the mean temporal average of the moving averaged filtered ADC signal and a subsequent FFT are carried out. Results show that there is a smaller sensor signal when the applied force or bending at the sensor is higher. In all investigations, there was a different sensor signal. The flex-bending sensor can detect the effect of pulsatile flow. The smallest temporal averaged signal difference between reference and clamp in the front wire's tip is 1.09%. For example, the mean temporal average of the filtered ADC signal for different clinically relevant scenarios is between 2550 and 2900. The results show that the sensorized guide wire developed for catheterization can support aortic valve implementation. The sensor sensitivity is sufficient to detect even very small variations within the blood vessel and therefore is promising to support catheterization heart valve surgeries in future.
Read full abstract