Antimicrobial susceptibility testing (AST) from blood culture (BC) may take several days, limiting the eventual impact on antimicrobial stewardship. Hence, rapid AST systems represent a valuable support in shorting the time-to-response. In this work, the Quantamatrix dRASTTM system (dRAST) was evaluated for rapid AST on 100 monomicrobial BCs (50 Gram-negatives and 50 Gram-positives), including several isolates with clinically relevant resistance mechanisms. AST results were provided in 6-hours, on average. Compared to Micronaut (Merlin) system based on broth microdilution, dRAST exhibited an overall categorical agreement of 92.5 %, essential agreement of 89.0 %, and mean bias of 15.9 %. Category overestimation (potentially leading to unnecessary high-dosage treatment or to exclude active agents) and category underestimation (potentially leading to underdosing or using ineffective agents) were observed in 4.3 % and 3.1 % of cases, respectively. Even though several issues were reported, results confirmed the potential contribution of dRAST to shorten the BCs clinical microbiology workflow and management.