IntroductionHaemodialysis patients who require defibrillator therapy are expected to benefit from the entirely avascular subcutaneous defibrillator (S-ICD), but haemodialysis is associated with dynamic changes in R and T wave amplitude which can impact S-ICD eligibility. A continuous assessment of S-ICD eligibility during haemodialysis has not previously been performed. Material and methodsContinuous surface ECG recordings were obtained from a cohort of patients undergoing maintenance haemodialysis, but without an indication for an ICD. Automated vector screening was retrospectively performed at one-minute intervals throughout the dialysis session. Variations in S-ICD eligibility were calculated and in vectors with high degrees of variation, the underlying mechanism was identified. Results72 vector recordings (mean duration 254.1 ± 6.0 min) were obtained from 24 patients (mean age 64.3 ± 5.5 years, 68% male). At the start of haemodialysis 47 vectors were S-ICD eligible (65.2%). At the end of session, all of these vectors had remained eligible, and an additional 6 vectors had also become eligible (73.6%). High vector score variability was observed in 7 patients and the commonest cause was a progressive change in R:T ratio (71.5%). ConclusionIn a haemodialysis population, a single haemodialysis session can be associated with a potential change in S-ICD eligibility in 8.4% of vectors, with up to 12.5% of vectors showing high degrees of variability, most commonly due to variations in R:T ratio. In an S-ICD population with similar characteristics S-ICD screening prior to haemodialysis would be expected to more accurately identify vectors that retain eligibility.