Abstract
The subcutaneous implantable cardioverter defibrillator (S-ICD) provides an attractive option for patients with congenital heart disease (CHD) in whom a transvenous defibrillator is contraindicated. Given the unusual cardiac anatomy and repolarization strain, the surface electrocardiogram (ECG) is frequently abnormal, potentially increasing the screen failure rate. We prospectively screened 100 adult CHD patients regardless of the presence of clinical indication for ICD utilizing a standard left sternal lead placement, as well as a right parasternal position. Baseline patient and 12-lead ECG characteristics were examined to assess for predictors of screen failure. Average patient age was 48±14 years, average QRS duration was 134±37 ms, and 13 patients were pacemaker dependent. Using the standard left parasternal electrode position, 21 patients failed screening. Of these 21 patients with screen failure, 9 passed screening with the use of right parasternal electrode positioning, reducing screening failure rate from 21% to 12%. QT interval and inverted T wave anywhere in V2-V6 leads were found to be independent predictors of left parasternal screening failure (P=0.01 and P=0.04, respectively). Utilization of both left and right parasternal screening should be used in evaluation of CHD patients for S-ICD eligibility. ECG repolarization characteristics were also identified as novel predictors of screening failure in this group. (Circ J 2016; 80: 1328-1335).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.