This study aimed to investigate the diagnostic value of computed tomography virtual intravascular endoscopy (VIE) in the follow-up of patients with abdominal aortic aneurysm (AAA) treated with fenestrated stent grafts. A total of 19 patients (17 males and 2 females; mean age: 75 years) with AAA undergoing fenestrated stent grafts were retrospectively studied. Pre- and post-fenestration computed tomography data were reconstructed for the generation of VIE images of aortic ostia and fenestrated stents and compared with two-dimensional axial and multiplanar reformation (MPR) images. Serum creatinine was measured pre and post fenestration to evaluate the renal function. The mean intra-aortic length measured by VIE, two-dimensional axial and MPR were 4.7, 4.4 and 4.6 mm, respectively, for the right renal stent; 5.0, 4.9 and 5.0 mm, respectively, for the left renal stent; and 5.9, 6.0 and 6.0 mm, respectively, for the superior mesenteric artery stent. Comparisons of these measurements did not show significant difference (P > 0.05). The mean diameters of renal artery ostia measured on VIE visualization pre and post fenestration were 9.2 x 8.3 and 10 x 8.9 mm for the right renal ostium; 8.3 x 7.1 and 9.9 x 8.9 mm for the left renal ostium, with significant changes observed (P < 0.01). No renal dysfunction was observed in this group. VIE is a valuable visualization tool in the follow-up of fenestrated stent graft repair of AAA by providing intraluminal appearance of fenestrated stents and measuring the length of stent protrusion.
Read full abstract