The purpose of this study was to develop a simplified method to approximate constants minimizing the standard deviation (SD) and the root mean square (RMS) of the prediction error in single-optimized intraocular lens (IOL) power calculation formulas. The study introduces analytical formulas to determine the optimal constant value for minimizing SD and RMS in single-optimized IOL power calculation formulas. These formulas were tested against various datasets containing biometric measurements from cataractous populations and included 10,330 eyes and 4 different IOL models. The study evaluated the effectiveness of the proposed method by comparing the outcomes with those obtained using traditional reference methods. In optimizing IOL constants, minor differences between reference and estimated A-constants were found, with the maximum deviation at -0.086 (SD, SRK/T, and Vivinex) and -0.003 (RMS, PEARL DGS, and Vivinex). The largest discrepancy for third-generation formulas was -0.027mm (SD, Haigis, and Vivinex) and 0.002mm (RMS, Hoffer Q, and PCB00/SN60WF). Maximum RMS differences were -0.021 and +0.021, both involving Hoffer Q. Post-minimization, the largest mean prediction error was 0.726 diopters (D; SD) and 0.043 D (RMS), with the highest SD and RMS after adjustments at 0.529 D and 0.875 D, respectively, indicating effective minimization strategies. The study simplifies the process of minimizing SD and RMS in single-optimized IOL power predictions, offering a valuable tool for clinicians. However, it also underscores the complexity of achieving balanced optimization and suggests the need for further research in this area. The study presents a novel, clinically practical approach for optimizing IOL power calculations.