Aseptic loosening (AL) is a frequent complication after rotating hinge knee (RHK) prosthesis. Citak's ratio has recently been developed to describe and classify distal femoral morphology into 3 groups (A, B, C). It consists in a ratio between the diameters of the femoral canal at 20cm from the knee joint line and at 2cm from the adductor tubercle. The objective of the study was to identify whether the femoral distal anatomical shape described with this ratio represents a risk factor for AL in RHK prosthesis. Retrospective study of patients who had undergone primary or revision RHK prosthesis, with a follow-up of minimum 4 years. Citak's ratio was calculated, and patients were classified depending on its value. Univariate and bivariate statistical analysis was performed to identify AL risk factors. Receiver Operating Characteristics (ROC) analyses were conducted to examine diagnostic quality of the parameters of interest. Thirty-three patients were included. Most of them females (ratio 26:7), with a mean age of 78.2 (SD 6.9). Three patients presented AL (rate of 9%), all of them classified into group C (100%). Citak's ratio was significantly related to the AL rate (p < 0.001), and so was the femoral canal diameter at 20cm from the knee joint (p 0.010). The ROC curve analysis yielded an Area Under the Curve (AUC) of 0.922 (CI 95% 0.819-1.000) for the Citak´s ratio. The inner femoral diameter at 20cm proximal to the knee joint line and Citak's ratio help indentify patients at risk of AL after RHK prosthesis, and thus a better planning of the surgery.