Size-exclusion chromatography with light scattering detection experiments conducted on poly(acrylic acid) neutralized to different degrees or using hydroxides with different counterions suggest that the same counterion and degree of neutralization is observed at the detector, irrespective of salt concentration, degree of neutralization and counterion at the time of injection. This strongly supports that during the chromatographic experiment the counterions of the polyelectrolyte are exchanged with those of the eluent, resulting in an effective dialysis of the polyelectrolyte solution during the size-exclusion chromatography experiment. Consequently, the refractive index increment determined by a refractive index detector equals the refractive index increment obtained after excessive dialysis against the pure eluent. Therefore, the species detected and characterized by light scattering coupled to size-exclusion chromatography are not identical to the species injected into the chromatographic system. Despite this structural change during the chromatographic experiments, the correct molar mass for the injected species is obtained by size-exclusion chromatography with light scattering detection.