ObjectiveLeft subclavian artery (LSA)-branched endografts with retrograde inner branch configuration (thoracic branch endoprosthesis [TBE]) offer a complete endovascular solution when LSA preservation is required during zone 2 thoracic endovascular aortic repair. However, the hemodynamic consequences of the TBE have not been well-investigated. We compared near-wall hemodynamic parameters before and after the TBE implantation using computational fluid dynamic simulations. MethodsEleven patients who had undergone TBE implantation were included. Three-dimensional aortic arch geometries were constructed from the pre- and post-TBE implantation computed tomography images. The resulting 22 three-dimensional aortic arch geometries were then discretized into finite element meshes for computational fluid dynamic simulations. Inflow boundary conditions were prescribed using normal physiological pulsatile circulation. Outlet boundary conditions consisted of Windkessel models with previously published values. Blood flow, modeled as Newtonian fluid, simulations were performed with rigid wall assumptions using SimVascular's incompressible Navier-Stokes solver. We compared well-established hemodynamic descriptors: pressure, flow rate, time-averaged wall shear stress (TAWSS), the oscillatory shear index (OSI), and percent area with an OSI of >0.2. Data were presented on the stented portion of the LSA. ResultsTBE implantation was associated with a small decrease in peak LSA pressure (153 mm Hg; interquartile range [IQR], 151-154 mm Hg vs 159 mm Hg; IQR, 158-160 mm Hg; P = .005). No difference was observed in peak LSA flow rates before and after implantation: 40.4 cm3/ (IQR, 39.5-41.6 cm3/s) vs 41.3 cm3/s (IQR, 37.2-44.8 cm3/s; P = .59). There was a significant postimplantation increase in TAWSS (15.2 dynes/cm2 [IQR, 12.2-17.7 dynes/cm2] vs 6.2 dynes/cm2 [IQR, 5.7-10.3 dynes/cm2]; P = .003), leading to decreases in both the OSI (0.088 [IQR, 0.063 to –0.099] vs 0.1 [IQR, 0.096-0.16]; P = .03) and percentage of area with an OSI of >0.2 (10.4 [IQR, 5.8-15.8] vs 15.7 [IQR, 10.7-31.9]; P = .13). Neither LSA side branch angulation (median, 81°, IQR, 77°-109°) nor moderate compression (16%-58%) seemed to have an impact on the pressure, flow rate, TAWSS, or percentage of area with an OSI of >0.2 in the stented LSA. ConclusionsThe implantation of TBE produces modest hemodynamic disturbances that are unlikely to result in clinically relevant changes.