BackgroundModular metaphyseal engaging (MME) femoral components in total hip arthroplasty (THA) allow optimized femoral length, offset, and anteversion and are useful in patients with unusual proximal femoral anatomy. Fretting, corrosion, and stem fractures above the modular sleeve are complications associated with these implants. The purpose of this study was to identify failure mechanisms of retrieved MME femoral components at our institution, identify all broken stem cases, and evaluate how often an extended trochanteric osteotomy (ETO) was required for removal. MethodsAll consecutively retrieved MME femoral components from September 2002 to May 2023 were reviewed. Patient demographics, procedure information, component specifications, indications for removal, and requirements for further revision surgery were reviewed. Descriptive statistics were calculated for variables of interest. ResultsThere were 131 retrieved MME components. The mean age at surgery was 59 years (range, 28 to 75), 49% were women, mean body mass index was 29.4 (range, 20.7 to 33.3), and mean American Society of Anesthesiologists score was 2.4 ± 0.5. There were 102 (78%) stems of 1 design (stem A), and 29 (22%) stems of a different design (stem B). Of 131 components, 10 (7.6%) failed secondary to stem fracture proximal to the modular sleeve. Four of 102 (4%) of stem A and 6 of 29 (21% of stem B) fractured. All broken stems required additional intervention for removal during revision THA, using an ETO (N = 9) or cortical window (N = 1) in which an intraoperative proximal femoral fracture occurred. ConclusionsBroken MME stems present a challenge for orthopaedic surgeons during revision THA. When a stem fracture occurs above the ingrown sleeve, the distal splines may have osseous interdigitation into the clothespin. Thus, when revising a broken MME stem, an ETO should be performed, and the segment should be long enough to allow distal access.