<h3>Purpose/Objective(s)</h3> In this work, we describe the technical aspects of our VMAT TBI technique, compare it to other VMAT TBI techniques, and share our initial experience. <h3>Materials/Methods</h3> From 2019 to 2021, 35 patients were treated with VMAT TBI at our institution. Treatment planning was performed using in-house developed automated planning scripts. Organ sparing depended on the regimen: myeloablative (lungs, kidneys, and lenses); non-myeloablative (lungs, kidneys, lenses, gonads, brain, and thyroid). Quality assurance was performed using EPID portal dosimetry and patient quality assurance platform. Robustness was evaluated for the first ten patients by performing local and global isocenter shifts of 5 mm. Treatment was delivered using IGRT for every isocenter and every fraction. In- vivo measurements were performed on the matchline between the VMAT and AP/PA fields and on the testes for the first fraction. <h3>Results</h3> The lungs, lungs-1cm, and kidneys D<sub>mean</sub> were consistently spared to 57.6<i>±</i>4.4%, 40.7<i>±</i>5.5%, and 70.0<i>±</i>9.9% of the prescription dose, respectively. Gonadal sparing (D<sub>mean</sub>=0.69<i>±</i>0.13 Gy) was achieved for all patients with benign disease. The average PTV D<sub>1cc</sub> was 120.7<i>±</i>6.4% for all patients. The average Gamma passing rate for the VMAT plans was 98.1<i>±</i>1.6% (criteria of 3%/2mm). Minimal differences were observed between Mobius3D- and EclipseAAA-calculated PTV D<sub>mean</sub> (0.0<i>±</i>0.3%) and lungs D<sub>mean</sub> (-2.5<i>±</i>1.2%). Robustness evaluation showed that the PTV D<sub>max</sub> and lungs D<sub>mean</sub> are insensitive to small positioning deviations between the VMAT isocenters (1.1<i>±</i>2.4% and 1.2<i>±</i>1.0%, respectively). The average matchline dose measurement indicated patient setup was reproducible (96.1<i>±</i>4.5% relative to prescription dose). Treatment time, including patient setup and beam-on, was 47.5±9.5 min. <h3>Conclusion</h3> VMAT TBI is a modern alternative to conventional 2D TBI treatment. Together with publicly shared autoplanning scripts, our technique may provide the gateway for wider adaptation of this technology and the possibility of multi-institutional studies in the cooperative group setting.
Read full abstract