A MWPC camera for nuclear medicine imaging with positron emitting isotopes has been built. The two planar detectors are converter coupled MWPC with sensitive area 48×48 cm 2 placed 50 cm apart with clear space in between for the patient. The effective sensitivity of 5.8% for each detector with a spatial resolution of 7 mm fwhm on the image plane is achieved by use of honeycomb lead-on-plastic converters. Present performance in imaging test objects and in clinical applications is described, using various computer 3D reconstruction algorithms. Developments in progress include the use of small diameter lead oxide glass tube converters; techniques to eliminate multiple scatter events and improve the point response function; and electronics to provide accidental event subtraction.