Purpose To characterize the morphology, size, and change in size of the contact zone of piggyback intraocular lenses (IOLs) of different materials and optic designs. Setting Department of Ophthalmology, Vienna General Hospital, Vienna, Austria. Methods In a prospective study, 9 eyes of 7 patients received piggyback IOLs of the following materials: poly(methyl methacrylate) (PMMA), acrylic, hydrogel, and silicone. The contact zone between the anterior and posterior IOLs was photodocumented from 1 day to 1 year after surgery using specular microscopy. The contact zone area was measured. Results A contact zone was present with all IOL materials studied. The area of contact, however, differed significantly. With PMMA IOLs, the contact zone was small and surrounded by Newton rings, indicating the tiny gap between the IOLs. With IOLs of soft material, such as silicone and hydrogel, it was larger than with PMMA IOLs and had a slightly irregular shape. With foldable acrylic IOLs, it was regular, round, and slightly larger than with the soft materials. The contact area enlarged primarily during the first 3 months after surgery. After 1 year, 2 eyes with acrylic piggyback IOLs had a membrane formation around the contact zone and 2 eyes developed Elschnig pearls between the IOLs. Conclusion In piggyback IOL eyes, the shape and size of the contact zone were strongly dependent on the IOL material and optic design. Contact area enlargement seemed to be induced by capsule shrinkage. Fibrous membrane formation around the contact zone and Elschnig pearl formation between the piggyback IOLs were long-term complications of this technique.
Read full abstract