BackgroundDespite growing clinical use of left bundle branch pacing (LBBP) there is scarcity of data regarding fundamentals of this pacing modality including chronaxie and rheobase. ObjectiveThe aims of this study were to calculate strength-duration curves with chronaxie, rheobase values for LBBP and left ventricular septal myocardial pacing (LVSP), to analyse battery current drain and presence of selective LBBP at very short pulse duration (PD). MethodsThe group of 141 patients with permanent LBBP were studied. The LBBP and LVSP capture thresholds were assessed at 6 different PDs to calculate the strength-duration curves. Battery current drain at these PDs and presence of selective LBBP were determined. For comparison of strength-duration curves between His bundle pacing (HBP) and LBBP, source data from our previous work based on 127 patients with HBP were obtained. ResultsThe chronaxies for LBBP and LVSP were very similar (0.38 ms vs. 0.39 ms) and the rheobases were identical (0.27V). The chronaxie for LBBP was lower than for HBP (0.38ms vs. 0.53 ms, p < 0.001), whereas rheobases were similar (0.27V vs. 0.26V). A narrow zone of selective capture was present in 19% and 41% of patients at PD of 0.06 ms and 0.03 ms, respectively. When pacing with the safety margin of +1 V, the lowest battery current drain was achieved with PD of 0.2 ms. ConclusionThe obtained strength-duration curves for LBBP and LVSP provide insights for optimal programming of LBBAP devices with regard to pulse duration, voltage amplitude, battery longevity and selective capture.