Aim:The purpose of this study was to evaluate the effects of three electronic apex locators (EAL), electric pulp tester (EPT) and diathermy on pacemaker function in vitro.Materials and Methods:Three EALs: Root ZX (J. Morita Co., Tustin, CA, U.S.A.), Propex (Dentsply), Mini Apex locator (SybronEndo, Anaheim, CA, USA), EPT (Parkell pulp vitality tester Farmingdale, NY, USA) and Diathermy (Neomed 250 B) were tested for any interference with one pacemaker (A medtronic kappa KVDD901-serial number: PLE734632S). Directly connecting the pacemaker lead with the EAL/EPT/diathermy operating on a flat bench top, the telemetry wand was held directly over the pacemaker to monitor the pacing pattern for a period of 30 s. Pacemaker activity was continuously recorded on the telemetric programmer and electro gram (EGM) readings examined for pacer inhibition, noise reversion or inappropriate pacemaker pulses.Results:All the three apex locators showed no pacing interference or background noise during its function or at rest. The EGM readings of EPT showed varying levels of background noise in between pacing however, this did not affect the normal pacing pattern and the pacing interval remained constant. EGM readings of diathermy showed an increase in the pacing interval (irregular pacing pattern) followed by complete inhibition of the pacing system.Conclusion:The tested EALs do not interfere with cardiac pacemaker function. The tested EPT showed varying levels of background noise but does not interfere with cardiac pacemaker function. Use of Diathermy interfered with the normal pacing, leading to complete inhibition of the pacing system.