We describe a multipurpose neutron powder diffractometer for high-efficiency and high-resolution measurements, HERMES, which has been installed at the JRR-3M reactor in Japan Atomic Energy Research Institute. To obtain high-quality data for a short counting time, or with a small sample amount, HERMES has a multi-detector system which consists of 150 3He neutron detectors and simple window type collimation. The best resolution is Δ2θ of 18' and Δd/d of 3 ×10-3, where 2θ and d are scattering-angle and d-spacing, respectively. The duration of an ordinary powder diffraction experiment is 1–5 h under the present conditions in JRR-3M. Some typical results of powder diffraction experiments and performance are given in this paper. We also show a result of 2D intensity mapping of magnetic diffuse scattering from a single crystal of a spin glass material.
Read full abstract