Objective: To compare the difference of decentration and tilt among 4 kinds of aspherical intraocular lenses (IOLs), and to analyze their objective visual quality. Methods: Prospective non-randomized controlled clinical trial. Age-related cataract patients who planned to undergo phacoemulsification and IOL implantation in the Affiliated Hospital of North Sichuan Medical College from April to June in 2020 were divided into ZCB00 group, SN60WF group, A1-UV group, and AO group according to IOL types. Thirty right eyes and thirty left eyes were selected in each group. Before operation and at 1 day, 1 week and 1 month postoperatively, decentration and tilt values were measured by a new swept-source anterior-segment optical coherence tomography device (CASIA2), and wavefront aberrations, objective scatter index (OSI), modulation transfer function cut off frequency (MTF cut off) and Strehl ratio (SR) were also examined. Values of decentration, tilt and visual quality compared among 4 groups were took from the right eye. One-way analysis of variance was used for inter-group comparison, and repeated measurement one-way analysis of variance was used for intra-group comparison. Data between right eyes and left eyes from all the individuals were compared by independent sample t-test. Results: A total of 181 patients (240 eyes) were enrolled, and 169 patients (224 eyes) completed the follow-up (114 right eyes and 110 left eyes). There were 77 males and 92 females, with an age of (69±9) years. There was no significant difference in gender, age, axial length, decentration and tilt of crystalline lens and IOL power among 4 groups (all P>0.05). At 1 day, 1 week and 1 month postoperatively, there was significant difference in decentration value among 4 groups (F=7.11, 6.12, 4.66; all P<0.05). For further pairwise comparison, the decentration value of SN60WF group was higher than that of the other 3 groups at 1 day and 1 week postoperatively, and the decentration value of SN60WF group was (0.259±0.101) mm at 1 month postoperatively, which was higher than that of ZCB00 group (0.177±0.099) mm and AO group (0.163±0.122) mm, and the differences were statistically significant (using SNK-q test, both P<0.05). The IOL tilt value in the ZCB00 group, SN60WF group, A1-UV group, and AO group at 1 month postoperatively were (4.806±1.129)°, (5.080±1.309)°, (4.586±1.338)°, (5.112±1.406)°, respectively. No significant difference in tilt value among 4 groups was found at any time after surgery (all P>0.05). In each group, there was no significant difference in decentration and tilt value at different postoperative time points (all P>0.05). At 1 month postoperatively, there was no significant difference in decentration and tilt at horizontal and vertical directions respectively among 4 groups (all P>0.05), and in each group, there was no significant difference in decentration and tilt value between right eyes and left eyes (all P>0.05), and IOLs tended to tilt towards the inferonasal or inferotemporal direction in both eyes. With 4-mm and 6-mm pupil diameter, there was significant difference in internal (F=131.32, 85.17) and ocular (F=46.64, 47.55) spherical aberration among 4 groups (all P<0.01). For further pairwise comparison, the spherical aberration of AO group was higher than that of the other 3 groups, and the difference was statistically significant (using SNK-q test, all P<0.05). There was no significant difference in OSI, MTF cut off and SR among 4 groups (all P>0.05). Conclusions: The four types of IOLs show decentration and tilt in varying degrees after implantation in the capsular bag, but this difference do not lead to clinical significance. Human eyes have tolerance to mild decentration and tilt of aspheric IOLs, showing good visual quality.