Abstract

BackgroundTo assess the precision and agreement of measurements of higher order aberrations (HOAs) obtained with a ray tracing aberrometer (iTrace) and a Hartmann-Shack aberrometer (Topcon KR-1 W).MethodsProspective evaluation of the diagnostic test. Data from the right eyes of 92 normal subjects obtained using the two devices were included in this study. Two observers performed 3 consecutive scans to determine the intraobserver repeatability and interobserver reproducibility. About one week later, one observer performed an additional 3 consecutive scans to obtain the intersession reproducibility. The within-subject standard deviation (Sw), test-retest repeatability (TRT) and intraclass correlation coefficient (ICC) were used to assess the precision, while Bland-Altman plots were performed to assess the agreement.ResultsFor intraobserver repeatability of the ocular, corneal and internal HOAs, Topcon KR-1 W showed a 2.77Sw of 0.079 μm or less and ICCs of 0.761 or more; and iTrace showed a 2.77Sw of 0.105 μm or less and ICCs of 0.805 or more. The ICCs of the internal HOAs of interobserver reproducibility were less than 0.75 except for spherical aberration (SA) (0.862), and interobserver reproducibility of the counterpart showed similar but lower results. For the ocular, corneal and internal HOA measurements, statistically significant differences existed between the Topcon KR-1 W and iTrace (all P < 0.05). No significant differences were observed in the ocular SA and internal coma.ConclusionsThe ray tracing and Hartmann-Shack method aberrometers provided excellent repeatability but less reliable reproducibility in the measurement of HOAs (except for SA). The two aberrometers should not be interchangeable in clinical application because of the significant differences in HOA measurements between them.

Highlights

  • To assess the precision and agreement of measurements of higher order aberrations (HOAs) obtained with a ray tracing aberrometer and a Hartmann-Shack aberrometer (Topcon KR-1 W)

  • We evaluated the repeatability and reproducibility of corneal power measurements obtained by Topcon KR-1 W and iTrace [15]

  • Since the more important function of the two devices is the measurement of HOAs, in this study, we estimated the precision and agreement of ocular, corneal and internal HOAs under 4 mm pupil diameter obtained by Topcon KR-1 W and iTrace in normal eyes

Read more

Summary

Introduction

To assess the precision and agreement of measurements of higher order aberrations (HOAs) obtained with a ray tracing aberrometer (iTrace) and a Hartmann-Shack aberrometer (Topcon KR-1 W). To the best of our knowledge, few studies have reported the assessment of precision (repeatability and reproducibility) and agreement of HOAs obtained by the two devices, simultaneously. We evaluated the repeatability and reproducibility of corneal power measurements obtained by Topcon KR-1 W and iTrace [15]. Since the more important function of the two devices is the measurement of HOAs, in this study, we estimated the precision (repeatability and reproducibility) and agreement of ocular, corneal and internal HOAs under 4 mm pupil diameter obtained by Topcon KR-1 W and iTrace in normal eyes

Methods
Results
Conclusion