In this study, we aimed to evaluate dosimetric quality of total marrow and lymphoid irradiation (TMLI) plans for a ring gantry-based medical Linac with a two-layer multi-leaf collimator. We retrospectively retrieved treatment planning CT images, structure sets, and plan dose for four adult patients, two male and two female, who previously received TMLI treatments on helical tomotherapy (HT) at our institution. TMLI plans were optimized for a ring gantry-based medical Linac with a two-layer multi-leaf collimator (Halcyon, Varian Medical Systems, Inc., Palo Alto, CA). A prescription dose of 12 Gy in 8 fractions was prescribed to the skeletal bones from the skull to mid-thigh, spleen, spinal canal, and lymphoid volume. Five or six isocenters were placed with equal spacing along the patient's longitudinal direction in each TMLI plan with two 6-MV flattening filter-free volumetric modulated arc therapy (VMAT) fields at each isocenter. Isocenter separation ranged from 15 cm to 16.5 cm. Each VMAT field has a field size of 28 cm to 28 cm with the collimator at 90° and a full gantry rotation. The nominal dose rate was 800 MU/minute, and the maximum gantry rotation speed was 24°/sec. Institutional dosimetric constraints were used for optimization including a mean lung dose limit of less than 8 Gy. All the plans were normalized so that 85% the primary planning target volume received the prescription dose. The average mean doses to the target volumes ranged from 12.2 to 12.6 Gy in the Halcyon TMLI plans, while they ranged from 12.1 to 12.5 Gy in the HT TMLI plans. Relative to the prescription dose, the average mean dose for normal organs ranged from 21.3% to 56.6% in the Halcyon TMLI plans, while it ranged from 10.1% to 68.4% in the clinical HT plans. The difference in the average mean dose to normal organs was less than 0.5 Gy except two organs between the Halcyon and HT TMLI plans. The average median dose for normal organs ranged from 18.2% to 48.8% relative to the prescription dose in the Halcyon TMLI plans. The mean lung dose (MLD) in the Halcyon TMLI plans met the institutional limit with an average dose of 6.75±0.42 Gy (range: 6.44 - 7.36 Gy), while the average MLD was 6.54±0.77 Gy (range: 6.24 - 7.22 Gy) in the HT plans (p-value = 0.71 in the paired t-test). The average total monitor unit in the Halcyon TMLI plans was 4,425±906 MU (range: 3,470 - 5,575 MU) with an average beam-on time of 5.1±1.3 minutes (range: 4.1 - 7.0 minutes), which excludes isocenter setup time, while the average beam-on time was 22.2±3.2 minutes (range: 19.6 - 26.1 minutes) with the HT plans. Halcyon TMLI plans met our institutional dosimetric constraints with adequate normal organ sparing and target dose coverage. The beam-on time with the Halcyon plans was significantly shorter than that with the HT plans, which could lead to shorter treatment time and increased patient comfort. This study showed the feasibility of TMLI treatments on the Halcyon machine. The same method could be used for total body irradiation on Halcyon.