Correct inflation of the cuff of any breathing tube is important; overinflation can damage the mucosa of the tracheobronchial tree and underinflation will cause a leak. Three different techniques to determine cuff seal/leak during inflation of the bronchial cuff of left double-lumen endobronchial tubes (DLT) were each evaluated in 10 patients. DLT size was chosen from a formula based on the patient's height and sex. In the first technique, designated the positive pressure technique of bronchial cuff inflation (PPT), the bronchial side of the DLT is pressurized during connection of the tracheal side to a beaker of water. Air bubbles will appear in the beaker in the absence of bronchial seal. The second technique, designated the CO2 analysis technique of bronchial cuff inflation (CAT), is based on analysis of CO2 content of gas sampled from the tracheal side of the DLT during ventilation of its bronchial side. When the bronchus is sealed, the normal CO2 waveform changes to a flat line. In the third technique, designated the negative pressure technique of bronchial cuff inflation (NPT), suction is applied to the tracheal side of the DLT. The absence of bronchial seal will result in collapse of the reservoir bag within a breathing system connected to the bronchial side of the tube. The bronchial sealing volumes were 1.1 +/- 0.9 mL, 0.9 +/- 0.7 mL, and 0.3 +/- 0.4 mL (mean +/- SD) when measured with the PPT, CAT, and NPT, respectively.(ABSTRACT TRUNCATED AT 250 WORDS)