PurposeDose to the rectum during brachytherapy treatment may differ from an approved treatment plan which can be quantified with in vivo dosimetry (IVD). This study compares the planned with in vivo doses measured with MOSkin and PTW 9112 rectal probe in patients undergoing CT based HDR cervical brachytherapy with Co-60 source. MethodsDose measurement of a standard pear-shaped plan carried out in phantom to verify the MOSkin dose measurement accuracy. With MOSkin attached to the third diode, RP3 of the PTW 9112, both detectors were inserted into patients’ rectum. The RP3 and MOSkin measured doses in 18 sessions as well as the maximum measured doses from PTW 9112, RPmax in 48 sessions were compared to the planned doses. ResultsPercentage dose differences ΔD (%) in phantom study for two MOSkin found to be 2.22 ± 0.07% and 2.5 ± 0.07%. IVD of 18 sessions resulted in ΔD(%) of −16.3% to 14.9% with MOSkin and ΔD(%) of −35.7% to −2.1% with RP3. In 48 sessions, RPmax recorded ΔD(%) of −37.1% to 11.0%. MOSkin_measured doses were higher in 44.4% (8/18) sessions, while RP3_measured were lower than planned doses in all sessions. RPmax_measured were lower in 87.5% of applications (42/47). ConclusionsThe delivered doses proven to deviate from planned doses due to unavoidable shift between imaging and treatment as measured with MOSkin and PTW 9112 detectors. The integration of MOSkin on commercial PTW 9112 surface found to be feasible for rectal dose IVD during cervical HDR ICBT.
Read full abstract