The objective of this study was to assess the clinical performances of PhenoMATRIX and PhenoMATRIX PLUS for the screening of methicillin-resistant Staphylococcus aureus (MRSA) from nasal and inguinal/perineal ESwabs using chromogenic media. The automated performances were compared to the manual reading. Additionally, we evaluated PhenoMATRIX PLUS for the automatic release of the negative results to the Laboratory Information System (LIS) and the automatic discharge of the negative plates from the incubators. A total of 6,771 non-duplicate specimens were used by PhenoMATRIX as a machine learning model. The validation of these settings was performed on 17,223 non-duplicate specimens. The MRSA positivity rate was 5% (866/17,223). Validated settings were then used by PhenoMATRIX PLUS on another 1,409 non-duplicate specimens. The sensitivities of PhenoMATRIX and PhenoMATRIX PLUS were 99.8% [95% confidence interval (CI), 99.2%-99.9%] and 100% (95% CI, 92.1%-100%), respectively. The specificities of PhenoMATRIX and PhenoMATRIX PLUS were 99.1% (95% CI, 99.0%-99.2%) and 95.2% (95% CI, 93.8%-96.1%), respectively. All the 1,297 MRSA-negative specimens analyzed by PhenoMATRIX PLUS were automatically released and sent to the LIS immediately after availability of the culture image on the WASPLab (100% accuracy). All negative media plates were automatically discarded. PhenoMATRIX PLUS decreases the time spent by technologists on negative plates and ensures optimal usage of the incubators' capacity.