This study aimed to evaluate the quality of locally advanced nasopharyngeal carcinoma (NPC) radiotherapy plans generated by the automated planning module of acommercial treatment planning system (TPS). Data of 30patients with locally advanced NPC were retrospectively investigated. For each patient, volumetric modulated arc therapy (VMAT) plans with double arcs were generated manually by experienced physicists and automatically in the Pinnacle3 Auto-Planning module (Philips Medical Systems, Fitchburg, WI, USA). The anatomic distance between the second clinical target volume (CTV2) and the pons of the brainstem, and the Tcategory of disease were factored into the evaluation. Dosimetric verification was evaluated in terms of gamma pass rate. Target coverage, sparing of organs at risk (OARs), and monitor units were evaluated and compared between the manual and automatic VMAT plans. Not all treatment plans fully met the dose objectives for planning target volumes (PTVs) and OARs, particularly in T4 patients. Overall, automatic VMAT provides acomparable or superior plan quality to manual VMAT in most cases. In stratified analysis, plan quality is mainly independent on Tcategory but is also affected by anatomic distance. If the anatomic distance is less than 5 mm, the automatic VMAT plan quality is equal or even inferior to manual VMAT performed by experienced physicists. Conversely, if the anatomic distance is greater than 5 mm, the automatic VMAT plan quality is superior to manual VMAT. Gamma pass rates for quality assurance are similar between manual and automatic VMAT plans for the former case, but significantly higher in automatic VMAT for the latter. The selection of manual versus automatic VMAT planning in locally advanced NPC should be made individually based on the anatomic distance, rather than blindly and habitually, since automatic VMAT is not good enough to completely replace manual VMAT.