BackgroundConduction system pacing (CSP), including His-bundle pacing (HBP) and left bundle branch area pacing (LBBAP), has been used as an alternative for pacemaker indicated patients requiring ventricular pacing. ObjectiveThe purpose of this study was to characterize the safety and performance of HBP and LBBAP among patients enrolled in the Medtronic product surveillance registry (PSR). MethodsThis observational analysis included patients who underwent pacemaker implantations for HBP or LBBAP with a Medtronic Model 3830 lead between January 2019 and December 2023 in the Medtronic PSR. The primary outcomes were lead-related complications and pacing capture threshold. Baseline characteristics, R-wave amplitude, impedance, and all-cause mortality were summarized. ResultsA total of 2342 patients were included across 77 centers (mean age 74 years; 38.9% female). Of the patients implanted with a 3830 lead for CSP, 64.1% (n = 1502) had LBBAP placement and 35.9% (n = 840) had HBP placement. The most commonly reported indications for CSP were sinus node dysfunction (67.0%) and atrioventricular block (57.2%). LBBAP had lower pacing thresholds, higher R-wave sensing, and higher impedance (all P <.001) through 30 months. At 36 months postimplant, the lead complication rate for LBBAP and HBP was 2.5% and 6.3%, ,respectively with no difference in all-cause mortality. ConclusionIn a multicenter cohort of LBBAP and HBP patients treated with the catheter-delivered 3830 lead, lead-related complication rates were low and electrical parameters were stable through 30 months.