PurposeThe purpose of this study was to compare the degree of perilymphatic enhancement between 4 hour post-contrast constant flip angle three-dimensional fluid attenuated inversion recovery (3D-FLAIR) images obtained with short repetition time (TR) and those obtained with long TR. Materials and methodsThis single-center, prospective study included patients who underwent MRI of the inner ear with heavily T2-weighted sequence, 3D-FLAIR sequence with a “short” TR of 10,000 ms (s3D-FLAIR) and with a “long” TR of 16,000 ms (l3D-FLAIR). Signal intensity ratio (SIR) and contrast-to-noise ratio (CNR) obtained with s3D-FLAIR and l3D-FLAIR were quantitatively assessed using region of interest (ROI) method and compared. The morphology of the endolymphatic space on both sequences was also evaluated. ResultsFrom March 2020 to July 2020, 20 consecutive patients were enrolled (9 women and 11 men; mean age, 52.1 ± 14.5 [SD] years; age range: 29–75 years). On l3D-FLAIR images, mean SIR (21.1 ± 8.8 [SD]; range: 7.6–46.1) was significantly greater than that on s3D-FLAIR images (15.7 ± 6.7 [SD]; range: 5.9–33.4) (P < 0.01). On l3D-FLAIR images, mean CNR (17 ± 8.5 [SD]; range: 2–40) was significantly greater than that on s3D-FLAIR images (12 ± 6.3 [SD]; range: 3.2–29.8) (P < 0.01). Kappa value for inter-rater agreement for endolymphatic hydrops, vestibular atelectasis and perilymphatic fistula were 0.93 (95% CI: 0.74–1), 1 (95% CI: 0.85–1) and 1 (95% CI: 0.85–1) respectively. ConclusionThis study demonstrates that the sensitivity of 3D-FLAIR sequences to low concentration gadolinium in the perilymphatic space is improved by elongation of the TR, with SIR and CNR increased by +34.4% and +41.3% respectively.