Purpose: Analysis of the performance of the microSTARii reader for optically stimulated luminescence dosimeters (OSLD) used by the IROC Houston Quality Assurance Center (IROC HQAC) for external audits compare to the characteristics of the Microstar reader. Methods: The IROC HQAC uses the Microstar reader for the OSLD program for verification of output of photon, electron and proton beams. The calculation of dose from the OSLD system is based on a group of factors defined at time of the commissioning of a batch of detectors. Factors like system sensitivity (SS), depletion (KD), element correction factor (ECF), linearity (KL), energy correction (KE). The new microSTARii unit presents many hardware and software upgrades that were considered useful for this program. Based on these changes many factors, that were considered reader dependent, were revised in order to analyze the effect of the new reading process. The SS, KD, ECF and KL were evaluated and compared with data defined based on reading done on Microstar reader. Results: The SS is reader specific and specified at 100 cGy dose level. This value is define per reading session and monitored over time. The KD factor was found to be different because of reading procedure are different. Themore » ECF values changed for a group of nano dots compare to values defined based on reading done on the Microstar reader. The KL was defined for the reader. Conclusion: The new microSTARii reader presents new features that improve the efficiency of the OSLD program at the IROC HQAC. New characterization is needed before final implementation is done to match the requirements of the existing OSLD system defined for the Microstar reader. Changes in uncertainty of the results has not been analized.« less