PurposeTo apply an independent GPU-accelerated Monte Carlo (MC) dose verification for CyberKnife M6 with Iris collimator and evaluate the dose calculation accuracy of RayTracing (TPS-RT) algorithm and Monte Carlo (TPS-MC) algorithm in the Precision treatment planning system (TPS).MethodsGPU-accelerated MC algorithm (ArcherQA-CK) was integrated into a commercial dose verification system, ArcherQA, to implement the patient-specific quality assurance in the CyberKnife M6 system. 30 clinical cases (10 cases in head, and 10 cases in chest, and 10 cases in abdomen) were collected in this study. For each case, three different dose calculation methods (TPS-MC, TPS-RT and ArcherQA-CK) were implemented based on the same treatment plan and compared with each other. For evaluation, the 3D global gamma analysis and dose parameters of the target volume and organs at risk (OARs) were analyzed comparatively.ResultsFor gamma pass rates at the criterion of 2%/2 mm, the results were over 98.0% for TPS-MC vs.TPS-RT, TPS-MC vs. ArcherQA-CK and TPS-RT vs. ArcherQA-CK in head cases, 84.9% for TPS-MC vs.TPS-RT, 98.0% for TPS-MC vs. ArcherQA-CK and 83.3% for TPS-RT vs. ArcherQA-CK in chest cases, 98.2% for TPS-MC vs.TPS-RT, 99.4% for TPS-MC vs. ArcherQA-CK and 94.5% for TPS-RT vs. ArcherQA-CK in abdomen cases. For dose parameters of planning target volume (PTV) in chest cases, the deviations of TPS-RT vs. TPS-MC and ArcherQA-CK vs. TPS-MC had significant difference (P < 0.01), and the deviations of TPS-RT vs. TPS-MC and TPS-RT vs. ArcherQA-CK were similar (P > 0.05). ArcherQA-CK had less calculation time compared with TPS-MC (1.66 min vs. 65.11 min).ConclusionsOur proposed MC dose engine (ArcherQA-CK) has a high degree of consistency with the Precision TPS-MC algorithm, which can quickly identify the calculation errors of TPS-RT algorithm for some chest cases. ArcherQA-CK can provide accurate patient-specific quality assurance in clinical practice.
Read full abstract