Occlusion of ventriculoperitoneal shunts placed after intraventricular hemorrhage occurs frequently. The objective of this study was to develop a hemorrhagic hydrocephalus model to assess the ability of an oscillating microactuator within the ventricular catheter (VC) to prevent shunt obstruction. An invitro hydrocephalus model with extreme risk of shunt obstruction was created. Phosphate-buffered saline, blood, and thrombin were driven through ventriculoperitoneal shunts for 8 hours. Five VCs were fitted with a microactuator and compared with 5 control VCs. The microactuator was actuated by an external magnetic field for 30 minutes. Pressure within the imitation lateral ventricle was measured. In the 5 control shunts, 6 obstructions developed (3 VC, 3 valve-distal catheter) compared with 1 obstruction (VC) in the 5 microactuator shunts. In the control and microactuator groups, the median volume exiting the shunts in 8 hours was 30 mL versus 256 mL. Median time to reach an intraventricular pressure of 40 mm Hg (13.8 minutes vs. >8 hours), median total time >40 mm Hg (6.2 hours vs. 0.0 hours), and median maximum pressure (192 mm Hg vs. 36 mm Hg) were significantly improved in the microactuator group (P < 0.01). In addition to protecting the VC, the microactuator appeared to prevent hematoma obstructing the valve or distal catheter, resulting in a much longer duration of low intraventricular pressures. A microactuator activated by placing the patient's head in an external magnetic field could reduce shunt obstructions in hemorrhagic hydrocephalus.