The aim of this study was to evaluate the distribution of new Corneal Visualisation Scheimpflug Technology (Corvis ST) parameters in normal, Post-laser in situ keratomileusis (LASIK), Post-LASIK keratectasia (KE) and keratoconus (KC) eyes, and explore the diagnostic ability of these parameters in distinguishing KE from LASIK eyes. Twenty-three normal eyes, 23 LASIK eyes, 23 KE eyes and 23 KC eyes were recruited in this study. The following new Corvis ST parameters were measured: Max Inverse Radius, deformation amplitude (DA) Ratio Max [2 mm], Pachy Slope, DA Ratio Max [1 mm], Ambrosio’s relational thickness horizontal (ARTh), Integrated Radius, stiffness parameter at first applanation (SP-A1) and Corvis biomechanical index (CBI). The general linear model, linear regression model, relation analysis and receiver operating characteristic (ROC) curve were performed. The Max Inverse Radius, DA Ratio Max [2 mm], Pachy Slope, DA Ratio Max [1 mm], Integrated Radius and CBI in LASIK eyes, KE eyes and KC eyes were higher than in normal eyes, while the ARTh and SP-A1 parameters were lower than in normal eyes. The KE eyes had higher Max Inverse Radius, DA Ratio Max [2 mm], Pachy Slope, DA Ratio Max [1 mm], Integrated Radius, and lower SP-A1 value than LASIK eyes (all P < 0.05). The central corneal thickness was related to the Pachy Slope (r = −0.485), ARTh (r = −0.766), SP-A1 (r = 0.618) in KE eyes (all P < 0.05). The area under the ROC curve of Integrated Radius, Max Inverse Radius, DA Ratio Max [2 mm] and SP-A1 were above 0.800 in identifying KE from LASIK eyes. Thus, the new Corvis ST parameters were different between LASIK and KE eyes, suggesting that they might be helpful in distinguishing KE eyes from LASIK eyes. However, a further multi-center and large sample study is necessary to confirm these findings.