Thromboembolism in blood vessels poses a serious risk of stroke, heart attack, and even sudden death if not properly managed. Sonothrombolysis combined with ultrasound contrast agents has emerged as a promising approach for the effective treatment of thromboembolism. Recent reports have highlighted the potential of intravascular sonothrombolysis as a safe and effective treatment modality for deep vein thrombosis (DVT). However, its efficiency has not been validated through in vivo testing of retracted clots. This study aimed to develop a miniaturized multidirectional transducer featuring two four-layer lead zirconate titanate (PZT-5A) stacks with an aperture size of 1.4 mm × 1.4 mm, enabling both forward- and side-looking treatment. Integrated into a custom two-lumen 10-French (Fr) catheter, the capability of this device for intravascular sonothrombolysis was validated both in vitro and in vivo. With low-dose tissue plasminogen activators and nanodroplets, the rotational multidirectional transducer reduced the retracted clot mass (800 mg) by an average of 52% within 30 min during in vitro testing. The lysis rate was significantly higher by 37% than that in a forward-viewing transducer without rotation. This improvement was particularly noteworthy in the treatment of retracted clots. Notably, a long-retracted clot (> 10 cm) was successfully treated within 40 min in vivo by creating a flow channel with a diameter > 4 mm in a porcine DVT model. In conclusion, these findings strongly suggest the potential of this technique for clinical applications in sonothrombolysis, offering a feasible solution for effectively treating thromboembolism, particularly in challenging cases involving retracted clots.