National Physical Laboratory, the National Metrology Institute (NMI) of India is maintaining Primary standards of pressure that cover several decades of pressure, starting from 3.0E-06 Pa to 1.0 GPa. Among which a recent addition is a Force Balanced Piston Gauge, the non-rotating piston type, having better resolution and zero stability compared to any other primary pressure standards commercially available in the range 1.0 Pa to 15.0 kPa (abs and gauge). The characterization of this FPG is done against Ultrasonic Interferometer Manometer (UIM), the National Primary pressure standard, working in the range 1.0 Pa to 130.0 kPa (abs and diff) and Air Piston Gauge (APG), a Transfer Pressure Standard, working in the range 6.5 kPa to 360 kPa (abs and gauge), in their overlapping pressure regions covering both absolute and gauge pressures. As NPL being one of the signatories to the CIPM MRA, the Calibration and Measurement Capabilities (CMC) of both the reference standards (UIM & APG), are Peer reviewed and notified in the Key Comparison Data Base (KCDB) of BIPM. The estimated mean effective area of the Piston Cylinder assembly of this FPG against UIM (980.457 mm2) and APG (980.463 mm2) are well within 4 ppm and 10 ppm agreement respectively, with the manufacturer's reported value (980.453 mm2). The expanded uncertainty of this FPG, Q(0.012 Pa, 0.0025% of reading), evaluated against UIM as reference standard, is well within the reported value of the manufacturer, Q(0.008 Pa, 0.003% of reading) at k = 2. The results of the characterization along with experimental setup & measurement conditions (for gauge and absolute pressure measurements), uncertainty budget preparation and evaluation of measurement uncertainty are discussed in detail in this paper.