Abstract

The Korea Research Institute of Standards and Science (KRISS) has three major vacuum systems: an ultrasonic interferometer manometer (UIM; Section II, Figs. 1 and 2) for a low vacuum, a static expansion system (SES; Section III, Figs. 3 and 4) for a medium vacuum, and an orifice-type dynamic expansion system (DES, Section IV, Figs. 5 and 6) for high and ultra-high vacuum systems. For each system, explicit measurement model equations with multiple variables are given. According to ISO standards, all of these system variable errors were used to calculate the expanded uncertainty (U). For each system, the expanded uncertainties (k = 1, confidence level = 95%) and relative expanded uncertainty (expanded uncertainty/generated pressure) levels are summarized in Table 4. Within the uncertainty limits, our bilateral and key comparisons [CCM.P-K4 (10 Pa to 1 kPa)] are extensive and in good agreement with those of other nations (Fig. 8 and Table 5).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.