BackgroundFollowing implantation/replacement of an implantable cardioverter-defibrillator, patients are legally subjected to variable lengths of driving restrictions based on the indication (1 and 3 months after primary and secondary prevention, respectively; 1 week after device replacement). AimTo assess the incidence of ventricular arrhythmia during the theoretical driving restriction period in a large cohort of patients. MethodsPatients who underwent implantable cardioverter-defibrillator implantation for primary or secondary prevention or device replacement between 2015 and 2021 were included retrospectively. The primary endpoint was the occurrence of ventricular arrhythmia during the theoretical driving restriction period, as defined by guidelines. ResultsA total of 914 patients were analysed, including 654 first implantations (438 and 216 for primary and secondary prevention, respectively) and 260 device replacements. The primary outcome occurred in 2/438 patients (0.004%) during the 1-month period following device implantation for primary prevention and in 25/216 patients (11.5%) during the 3-month period following device implantation for secondary prevention; it did not occur in the 1-week period following device replacement. The monthly calculated risk of harm remained below the accepted threshold of 0.005% for each group. ConclusionsPrimary prevention patients, such as those who have undergone device replacement, have a low risk of ventricular arrhythmia, which could lead to a reduction in their driving restriction period. Secondary prevention patients experienced a higher risk of recurrent ventricular arrhythmia, supporting the 3-month driving restriction period.