Abstract Many medical examinations involve ionizing radiation. Although the range of available dosimeters is rather wide, their linearity and chemical stability are limited. Recently, there has been a growing interest in new, improved dosimetric materials for emerging applications in medicine and other fields, such as sterilisation of consumer goods and medical instruments, irradiation of seeds, chemical agents and others. One of the classical dosimeters is carbon-doped alumina (Al2O3:C) – a well-established and widely used material for personal and industrial dosimeter with a range of great properties, such as high sensitivity, wide linearity range and relative ease of production and handling. However, the demand for reliable dosimeters in a high-dose range is still only partially fulfilled, and alumina doped with chromium ions (Al2O3:Cr) can be a promising candidate. In this study, we explored alumina doped with chromium porous microparticles synthesized with a sol-gel method as a possible high dose dosimeter and evaluated its thermostimulated luminescence signal, dose response with two irradiation sources and measured long-time fading. It was found that although the TSL signal was quite complex (consisting of two main peaks above room temperature) and the long-term fading was significant (around 50 % in the span of 30 days), with sufficient optimisation the material could be used as a high-dose dosimeter for X-ray and beta irradiation. Wide high dose linearity range, physical and chemical characteristics, as well as low production costs and ease of synthesis make chromium (III) doped alumina a compelling candidate for applicability in various medical and industry fields.
Read full abstract