Abstract

PurposeThe purpose of this paper is to present a radiochromic film dosimeter containing polyvinyl alcohol (PVA) matrix and various concentrations of methyl red (MR) dye for high dose measurements.Design/methodology/approachThe MR-PVA films were exposed to irradiation up to 60 kGy using 60Co source of gamma ray. The ultraviolet and visible regions (UV/VIS) spectrophotometry were used to examine the optical density of pre-and post-irradiated dosimeters at 424 nm.FindingsThe dose sensitivity of MR-PVA films increases significantly with increasing MR dye concentrations in the dose range of 5 to 60 kGy. The impact of relative humidity, irradiation temperature, dose rate and the stability of the films has been analyzed. The overall uncertainty of the MR-PVA film dosimeter is 6.12% (Double Standard-deviation, 95% confidence level).Practical implicationsIt was found that the MR-PVA films may be used as high dose dosimeter with an acceptable overall uncertainty in routine industrial radiation processing.Originality/valueThe color bleaching of irradiated MR-PVA films in terms of specific absorbance curves increases significantly with increasing absorbed dose up to 60 kGy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call