To date, there is no consensus on how to standardize the assessment of ototoxicity in serial measurements. For the diagnosis of damage to the cochlear amplifier, measurement methods are required that have the highest possible test-retest reliability and validity for detecting persistent damage. Estimated distortion-product thresholds (LEDPT) based on short-pulse distortion-product otoacoustic emission (DPOAE) level maps use individually optimal DPOAE stimulus levels and allow reliable quantitative estimation of cochlea-related hearing loss. Hearing thresholds were estimated objectively using LEDPT and subjectively using modified Békésy tracking audiometry (LTA). Recordings were performed seven times within threemonths at 14frequencies (f2 = 1-14 kHz) in 20ears (PTA4 (0.5-4 kHz) < 20 dB HL). Reconstruction of the DPOAE growth behavior as afunction of the stimulus levels L1, L2 was performed on the basis of 21DPOAE amplitudes. Anumerical fit of anonlinear mathematical function to the three-dimensional DPOAE growth function yielded LEDPT for each stimulus frequency. For the combined analysis, probability distributions of hearing thresholds (LTA, LEDPT), DPOAE levels (LDP), and combinations thereof were determined. LTA and LEDPT each exhibited atest-retest reliability with amedian of absolute differences (AD) of 3.2 dB and 3.3 dB, respectively. Combining LEDPT, LDP, and LTA into asingle parameter yielded asignificantly smaller median AD of 2.0 dB. It is expected that an analysis paradigm based on acombination of LEDPT, suprathreshold LDP, and fine-structure-reduced LTA would achieve higher test performance (sensitivity and specificity), allowing reliable detection of pathological or regenerative changes in the outer hair cells.