PurposeThe goal of this work is to investigate the field width dependence of the volumetric modulated arc therapy (VMAT) plan quality and to propose a half field method to irradiate large volumes effectively with VMAT.Materials and methodsWe compared four different VMAT methods; namely three full field (3ff), four full field (4ff), three half field (3hf), four half field (4hf). To evaluate the impact of the field width on VMAT plan quality, 12 different size PTVs were created in the virtual phantom and treatment plans generated for each PTV were compared. The effectiveness of our half field method was tested using computed tomography (CT) data of 10 nasopharyngeal carcinoma patients.ResultsIn the virtual phantom study, organs at risk (OAR) mean dose, the maximum point dose, and Homogeneity Index (HI) were found to be field width dependent. Conformation Number (CN) was not significantly affected. In the clinical study, 4hf plans obtained statistically significant dose reduction at brainstem (P < 0.001), right parotid (P = 0.034), oral cavity (P < 0.001), larynx (P = 0.003), cochlea (P = 0.017), lips (P = 0.024), and Body‐PTV (P = 0.04) compared to 4ff plans.ConclusionOur results indicate that VMAT plan quality is dependent on the field width. Half field VMAT method, with the help of reduced field width, shows a clear advantage for the irradiation of large size targets compared to traditionally used full field VMAT plans.