Background : For radiotherapy of total skin including sub-cutaneousssue up to a depth of 3 cm on the enre leleg of an adult (Angiosarcoma skin), a complex treatment with mulple staonary electron fi elds was planned at our clinic. The details of dosimetry, clinical dose measurements are presented. Materials and Methods : The treatment planned with 6 overlapping 9 MeV electron fields in Clinac 2300CD linac. With 25×25 cm cone, a cut-out insert provided 56 × 30 cm field at FSD 213 cm, while paent lying on the floor. Dose distribuons were checked using Kodak V densitometric film in cylindrical plasc can phantom. The calibraon was carried out using solid water phantom, water equivalent IMRT phantom and water can leg phantom. A dose of 45 Gy in 23 fracons at 5 fracons/week was prescribed. 6 field overlapping field factor was measured by the method described for total body electron irradiaon (AAPM). Skin doses were esmated at random selected points using TLD chips and semiconductor diodes. Results : Measured absorbed doses by three methods were 0.174 cGy/MU, 0.166 cGy/MU and 0.162 cGy/MU agreed well with the calculated value 0.163 cGy/MU. 6 field overlap factor was 2.315. Clinical dose esmates of mean skin dose was 246.0 + 14 cGy (n=18), delivering higher dose by 23%. The gonad dose esmate under shield was <5%. The excess dose to skin delivered in first 14 fracons was adjusted in following 9 fracons. Conclusion : It appears that the excess dose in real situaon may be due to either floor backscaDer or non uniform overlap of dose from adjacent fields.