Guidelines for bone scintigraphy are well established and recommend the use of planar early phase images to investigate a number of clinical indications. With recent advances in gamma camera technology the use of SPECT/CT imaging in the early phases is now possible, offering the potential of improved diagnostic confidence and prognostic value. To date little work has been carried out to optimize the acquisition of early phase bone images using SPECT/CT with most of the available studies acquiring SPECT images after the traditional planar images to allow comparison of the two techniques. Imaging durations of 7 to 10 minutes have been commonly used. However, the use of iterative reconstruction algorithms has been investigated with rapid SPECT imaging to allow imaging durations as low as 4 minutes. The use of CZT based systems with increased sensitivity and improved energy and spatial resolution also offers the potential to reduce imaging times. The optimization of projection measurement order has been investigated as a method of reducing image artefacts as a result of changing tracer distribution during the SPECT acquisition. In this article we consider the current state of early phase SPECT imaging and possible areas for future investigation as well as recommendations for departments looking to adopt blood pool SPECT imaging as part of their routine clinical practice.