ObjectivesEvaluation of the mechanical integrity and reliability of polylactide-based miniplates for osseous free flap fixation at the mandible in an experimental study setup of a mandible reconstruction model. Material and methods1.0mm titanium miniplates (group TI) (MatrixMandible, DePuy Synthes, Umkirch, Germany) and 1.5mm polylactide miniplates (group PL) (Inion CPS, Inion Oy, Tampere, Finland) were used to fix a polyurethane (PU) fibula segment to a PU mandible reconstruction model using monocortical non-locking screws. Mastication was simulated via unilateral cyclic dynamic loading at 1Hz with increasing loads (+ 0.15N/cycle, Bionix, MTS, USA). A 3D optical tracking system (Aramis, GOM, Braunschweig, Germany) was used to determine interosteotomy movements (IOM). ResultsIOM were higher in the polylactide group (distal: P=0.001, mesial: P=0.001). Differences in mean stiffness (titanium: 478±68N/mm; polylactide: 425±38N/mm, P=0.240) and mean force at a vertical displacement of 1.0mm (titanium: 201.6±87.1N; polylactide: 141.3±29.9N, P=0.159) were not significant. ConclusionsThe results of this study suggest that polylactide-based miniplates provide reduced mechanical integrity and higher interosteotomy movements in comparison to titanium miniplates in vitro. Indications for clinical use of polylactide-based miniplates in mandible reconstruction have to be placed critically. Future studies will focus on clinical complications of polylactide-based plates in risk patients.